mlsanityData to 14 Oct 2026

No-reference based automatic parameter optimization for iterative reconstruction using a novel search space aware crow search algorithm

arXiv 2604.06246 ↗First posted 6 Apr 2026cs.CVcited by 0

Not on the map. Why?

Abstract (arXiv)

Iterative reconstruction technique's ability to reduce radiation exposure by using fewer projections has attracted significant attention. However, these methods typically require a precise tuning of several hyperparameters, which can have a major impact on reconstruction quality. Manually setting these parameters is time-consuming and increases the workload for human operators. In this paper, we introduce a novel fully automatic parameter optimization framework that can be applied to a wide range of Cone-beam computed tomography (CBCT) iterative reconstruction algorithms to determine optimal parameters without requiring a reference reconstruction. The proposed method incorporates a modified crow search algorithm (CSA) featuring a superior set-dependent local search mechanism, a search-space-aware global search strategy, and an objective-driven balance between local and global search. Additionally, to ensure an effective initial population, we propose a chaotic diagonal linear uniform initialization scheme that accelerates algorithm convergence. The performance of the proposed framework was evaluated on three imaging machines and four real datasets, as well as three different iterative reconstruction methods with the highest number of tunable parameters, representing the most challenging senario. The results indicate that the proposed method could outperform manual settings and CSA, with an 4.19% improvement in average fitness and 4.89% and 3.82% improvements on CHILL@UK and RPI_AXIS, respectively, which are two benchmark no-reference learning-based quality metrics. In addition, the qualitative results clearly show the superiority of the proposed method by maintaining fine details sharply. The overall performance of the proposed framework across different comparison scenarios demonstrates its effectiveness and robustness across all cases.

Key facts. No citing papers in the map yet (data through 14 Oct 2026). Builds on 7 of its 64 references that are in the map. First posted on arXiv 6 Apr 2026.

Builds on

7 of 64 references.

References · 7

  1. 2018The Unreasonable Effectiveness of Deep Features as a Perceptual Metriccited by 11,177
  2. 2020Image Quality Assessment: Unifying Structure and Texture Similaritycited by 811
  3. 2024A study of why we need to reassess full reference image quality assessment with medical imagescited by 17
  4. 2024A study on the adequacy of common IQA measures for medical imagescited by 10
  5. 2017Intelligent Parameter Tuning in Optimization-based Iterative CT Reconstruction via Deep Reinforcement Learningcited by 7
  6. 2024TIGRE v3: Efficient and easy to use iterative computed tomographic reconstruction toolbox for real datasetscited by 5
  7. 2024Parameter choices in HaarPSI for IQA with medical imagescited by 2

Built on by

Not cited yet.