Dealing with Segmentation Errors in Needle Reconstruction for MRI-Guided Brachytherapy
Not on the map. Why?
Abstract (arXiv)
Brachytherapy involves bringing a radioactive source near tumor tissue using implanted needles. Image-guided brachytherapy planning requires amongst others, the reconstruction of the needles. Manually annotating these needles on patient images can be a challenging and time-consuming task for medical professionals. For automatic needle reconstruction, a two-stage pipeline is commonly adopted, comprising a segmentation stage followed by a post-processing stage. While deep learning models are effective for segmentation, their results often contain errors. No currently existing post-processing technique is robust to all possible segmentation errors. We therefore propose adaptations to existing post-processing techniques mainly aimed at dealing with segmentation errors and thereby improving the reconstruction accuracy. Experiments on a prostate cancer dataset, based on MRI scans annotated by medical professionals, demonstrate that our proposed adaptations can help to effectively manage segmentation errors, with the best adapted post-processing technique achieving median needle-tip and needle-bottom point localization errors of 1.07 (IQR ± 1.04) mm and 0.43 (IQR ± 0.46) mm, respectively, and median shaft error of 0.75 (IQR ± 0.69) mm with 0 false positive and 0 false negative needles on a test set of 261 needles.
Key facts. No citing papers in the map yet (data through 14 Oct 2026). Builds on 0 of its 10 references that are in the map. First posted on arXiv 25 Jul 2025.
Builds on
0 of 10 references.
No references here yet.
Built on by
Not cited yet.