Chronoamperometry with Room-Temperature Ionic Liquids: Sub-Second Inference Techniques
Not on the map. Why?
Abstract (arXiv)
Chronoamperometry (CA) is a fundamental electrochemical technique used for quantifying redox-active species. However, in room-temperature ionic liquids (RTILs), the high viscosity and slow mass transport often lead to extended measurement durations. This paper presents a novel mathematical regression approach that reduces CA measurement windows to under 1 second, significantly faster than previously reported methods, which typically require 1-4 seconds or longer. By applying an inference algorithm to the initial transient current response, this method accurately predicts steady-state electrochemical parameters without requiring additional hardware modifications. The approach is validated through comparison with standard chronoamperometric techniques and is demonstrated to maintain reasonable accuracy while dramatically reducing data acquisition time. The implications of this technique are explored in analytical chemistry, sensor technology, and battery science, where rapid electrochemical quantification is critical. Our technique is focused on enabling faster multiplexing of chronoamperometric measurements for rapid olfactory and electrochemical analysis.
Key facts. No citing papers in the map yet (data through 14 Oct 2026). Builds on 0 of its 10 references that are in the map. First posted on arXiv 5 Jun 2025.
Builds on
0 of 10 references.
No references here yet.
Built on by