Volume Interpolated Breath-Hold Examination (VIBE) MRI generates images suitable for water and fat signal composition estimation. While the two-point VIBE provides water-fat-separated images, the six-point VIBE allows estimation of the effective transversal relaxation rate R2* and the proton density fat fraction (PDFF), which are imaging markers for health and disease. Ambiguity during signal reconstruction can lead to water-fat swaps. This shortcoming challenges the application of VIBE-MRI for automated PDFF analyses of large-scale clinical data and of population studies. This study develops an automated pipeline to detect and correct water-fat swaps in non-contrast-enhanced VIBE images. Our three-step pipeline begins with training a segmentation network to classify volumes as "fat-like" or "water-like," using synthetic water-fat swaps generated by merging fat and water volumes with Perlin noise. Next, a denoising diffusion image-to-image network predicts water volumes as signal priors for correction. Finally, we integrate this prior into a physics-constrained model to recover accurate water and fat signals. Our approach achieves a < 1% error rate in water-fat swap detection for a 6-point VIBE. Notably, swaps disproportionately affect individuals in the Underweight and Class 3 Obesity BMI categories. Our correction algorithm ensures accurate solution selection in chemical phase MRIs, enabling reliable PDFF estimation. This forms a solid technical foundation for automated large-scale population imaging analysis.

Key facts. No citing papers in the map yet (data through 14 Oct 2026). First posted on arXiv 20 Feb 2025.

Builds on

0 of 0 references.

No references here yet.

Built on by

Not cited yet.

Similar, not cited

Close in wording, no citation either way

  1. 2026Do We Really Need Diffusion? A Fast U-Net for Paired Medical Image Translation
  2. 2024A Physics-based Generative Model to Synthesize Training Datasets for MRI-based Fat Quantification
  3. 2025Reverse Imaging for Wide-spectrum Generalization of Cardiac MRI Segmentation
  4. 2025Diffusion-Based Synthesis of 3D T1w MPRAGE Images from Multi-Echo GRE with Multi-Parametric MRI Integration
  5. 2022Physics-informed Deep Diffusion MRI Reconstruction with Synthetic Data: Break Training Data Bottleneck in Artificial Intelligence
  6. 2023DiffGEPCI: 3D MRI Synthesis from mGRE Signals using 2.5D Diffusion Model
  7. 2026Distortion-Corrected Diffusion MRI Using Rotated-View EPI and Joint Field-Map/Image Estimation with Gaussian Primitives
  8. 2026WFM: 3D Wavelet Flow Matching for Ultrafast Multi-Modal MRI Synthesis