Phase Aberration Robust Beamformer for Planewave US Using Self-Supervised Learning
Not on the map. Why?
Abstract (arXiv)
Ultrasound (US) is widely used for clinical imaging applications thanks to its real-time and non-invasive nature. However, its lesion detectability is often limited in many applications due to the phase aberration artefact caused by variations in the speed of sound (SoS) within body parts. To address this, here we propose a novel self-supervised 3D CNN that enables phase aberration robust plane-wave imaging. Instead of aiming at estimating the SoS distribution as in conventional methods, our approach is unique in that the network is trained in a self-supervised manner to robustly generate a high-quality image from various phase aberrated images by modeling the variation in the speed of sound as stochastic. Experimental results using real measurements from tissue-mimicking phantom and textitin vivo scans confirmed that the proposed method can significantly reduce the phase aberration artifacts and improve the visual quality of deep scans.
Key facts. Cited by 1 paper in the map (data through 13 Oct 2026). Builds on 4 of its 37 references that are in the map. First posted on arXiv 16 Feb 2022.
Builds on
4 of 37 references.
References · 4
- 2019Adaptive and Compressive Beamforming Using Deep Learning for Medical Ultrasoundcited by 12
- 2019Deep Learning-based Universal Beamformer for Ultrasound Imagingcited by 4
- 2019Ultrasound Aberration Correction based on Local Speed-of-Sound Map Estimationcited by 3
- 2019The SVD Beamformer: Physical Principles and Application to Ultrafast Adaptive Ultrasoundcited by 2
Built on by
Citing papers per year: 2023: 1.
1 paper.