mlsanityData to 7 Oct 2026

Automated Left Ventricle Dimension Measurement in 2D Cardiac Ultrasound via an Anatomically Meaningful CNN Approach

arXiv 1911.02448 ↗First posted 6 Nov 2019eess.IV cs.CVcited by 9

Not on the map. Why?

Abstract (arXiv)

Two-dimensional echocardiography (2DE) measurements of left ventricle (LV) dimensions are highly significant markers of several cardiovascular diseases. These measurements are often used in clinical care despite suffering from large variability between observers. This variability is due to the challenging nature of accurately finding the correct temporal and spatial location of measurement endpoints in ultrasound images. These images often contain fuzzy boundaries and varying reflection patterns between frames. In this work, we present a convolutional neural network (CNN) based approach to automate 2DE LV measurements. Treating the problem as a landmark detection problem, we propose a modified U-Net CNN architecture to generate heatmaps of likely coordinate locations. To improve the network performance we use anatomically meaningful heatmaps as labels and train with a multi-component loss function. Our network achieves 13.4%, 6%, and 10.8% mean percent error on intraventricular septum (IVS), LV internal dimension (LVID), and LV posterior wall (LVPW) measurements respectively. The design outperforms other networks and matches or approaches intra-analyser expert error.

Key facts. Cited by 9 papers in the map (data through 7 Oct 2026). First posted on arXiv 6 Nov 2019.

Builds on

No reference list yet.

Built on by

Cited by year

Citing papers per year: 2021: 4; 2022: 0; 2023: 1; 2024: 1; 2025: 3.

9 papers.