Multi-shot Echo Planar Imaging for accelerated Cartesian MR Fingerprinting: an alternative to conventional spiral MR Fingerprinting

arXiv 1906.08195 ↗First posted 19 Jun 2019physics.med-ph eess.IVcited by 2

Not on the map. Why?

Abstract (arXiv)

Purpose: To develop an accelerated Cartesian MRF implementation using a multi-shot EPI sequence for rapid simultaneous quantification of T1 and T2 parameters. Methods: The proposed Cartesian MRF method involved the acquisition of highly subsampled MR images using a 16-shot EPI readout. A linearly varying flip angle train was used for rapid, simultaneous T1 and T2 quantification. The accuracy of parametric map estimations were improved by using an iterative projection algorithm. The results were compared to a conventional spiral MRF implementation. The acquisition time per slice was 8s and this method was validated on a phantom and a healthy volunteer brain in vivo. Results: Joint T1 and T2 estimations using the 16-shot EPI readout are in good agreement with the spiral implementation using the same acquisition parameters (deviation less than 3% for T1 and less than 4% for T2) for the healthy volunteer brain. The T1 and T2 values also agree with the conventional values previously reported in the literature. The visual quality of the multi-parametric maps generated by the multi-shot EPI-MRF and spiral-MRF implementations were comparable. Conclusion: The multi-shot EPI-MRF method generated accurate quantitative multi-parametric maps similar to conventional Spiral - MRF. This multi-shot approach achieved provides an alternative for performing MRF using an accelerated Cartesian readout, thereby increasing the potential usability of MRF.

Key facts. Cited by 2 papers in the map (data through 7 Oct 2026). First posted on arXiv 19 Jun 2019.

Builds on

No reference list yet.

Built on by

Cited by year

Citing papers per year: 2022: 2.

2 papers.