mlsanityData to 14 Oct 2026

Model-based Catheter Segmentation in MRI-images

arXiv 1705.06712 ↗First posted 18 May 2017cs.CVcited by 4

Not on the map. Why?

Abstract (arXiv)

Accurate and reliable segmentation of catheters in MR-guided interventions remains a challenge, and a step of critical importance in clinical workflows. In this work, under reasonable assumptions, mechanical model based heuristics guide the segmentation process allows correct catheter identification rates greater than 98% (error 2.88 mm), and reduction in outliers to one-fourth compared to the state of the art. Given distal tips, searching towards the proximal ends of the catheters is guided by mechanical models that are estimated on a per-catheter basis. Their bending characteristics are used to constrain the image feature based candidate points. The final catheter trajectories are hybrid sequences of individual points, each derived from model and image features. We evaluate the method on a database of 10 patient MRI scans including 101 manually segmented catheters. The mean errors were 1.40 mm and the median errors were 1.05 mm. The number of outliers deviating more than 2 mm from the gold standard is 7, and the number of outliers deviating more than 3 mm from the gold standard is just 2.

Key facts. Cited by 4 papers in the map (data through 14 Oct 2026). First posted on arXiv 18 May 2017.

Builds on

No reference list yet.

Built on by

Cited by year (2026 to 14 Oct)

Citing papers per year: 2021: 1; 2022: 0; 2023: 0; 2024: 2; 2025: 0; 2026: 1.

4 papers.