# An Implant-to-Wearable IR-UWB Transmitter-Receiver Architecture and Layered Protocol for High-Density Brain-Computer Interfaces

arXiv: https://arxiv.org/abs/2609.22207 · HTML: https://mlsanity.com/paper/2609.22207/

## Key facts

- No citing papers in the map yet (data through 14 Oct 2026).
- Builds on 0 of its 11 references that are in the map.
- First posted on arXiv 1 Sep 2026.

## Abstract (arXiv)

High-rate neural telemetry requires an implant-towearable link whose circuit partition, waveform mapping, packet processing, and failure semantics are mutually consistent. This paper specifies a chip-oriented, one-way impulse-radio ultrawideband (IR-UWB) uplink comprising an implanted packet engine and free-running transmitter, a tissue-proxy channel, and a wearable mixed-signal receiver and digital baseband.