# Restoring the saturation response of a PMT using pulse-shape and artificial-neural-networks

arXiv: https://arxiv.org/abs/2302.06170 · HTML: https://mlsanity.com/paper/2302.06170/

## Key facts

- No citing papers in the map yet (data through 14 Oct 2026).
- Builds on 0 of its 4 references that are in the map.
- First posted on arXiv 13 Feb 2023.

## Abstract (arXiv)

The linear response of a photomultiplier tube (PMT) is a required property for photon counting and reconstruction of the neutrino energy. The linearity valid region and the saturation response of PMT were investigated using a linear-alkyl-benzene (LAB)-based liquid scintillator. A correlation was observed between the two different saturation responses, with pulse-shape distortion and pulse-area decrease. The observed pulse-shape provides useful information for the estimation of the linearity region relative to the pulse-area. This correlation-based diagnosis allows an in-situ estimation of the linearity range, which was previously challenging. The measured correlation between the two saturation responses was employed to train an artificial-neural-network (ANN) to predict the decrease in pulse-area from the observed pulse-shape. The ANN-predicted pulse-area decrease enables the prediction of the ideal number of photoelectrons irrelevant to the saturation behavior. This pulse-shape-based machine learning technique offers a novel method for restoring the saturation response of PMTs.